Key findings

  • 33nurses at nine hospitals in Denmark completed the 12-item survey
  • 90.9%rated the mechanical infusion system as easy or very easy to use (30/33)
  • 96.9%frequently or almost always had more time to interact and communicate with patients (31/32)
  • 81.3%reported less hand pain or discomfort than with manual push administration (26/32)

Abstract

Background
Subcutaneous administration of medicinal products is increasingly used in clinical practice. For short-duration administrations, manual push is a practical delivery method but requires continuous hands-on involvement by nurses. Mechanical pump-assisted administration may provide an alternative; however, evidence regarding nurses’ experiences with this approach for short-duration subcutaneous infusions is limited.
Methods
A descriptive, cross-sectional survey study was conducted among nurses at participating hospitals in Denmark to evaluate their experience with a non-electronic, spring-driven mechanical infusion system compared with routine manual push administration. The hospitals had independently elected to use the mechanical system in clinical practice. Following standard product training and practical use of the system, nurses completed a 12-item questionnaire assessing usability, setup time, troubleshooting, nurse–patient interaction, nurse hand discomfort, patient-reported pain as reported by nurses, and overall acceptance. Outcomes were summarized descriptively.
Results
Thirty-three nurses completed the survey. Overall, 90.9% rated the mechanical system as easy or very easy to use, and 97.0% reported feeling comfortable or very comfortable operating it. Setup time was reported as the same or shorter than manual push by 90.9% of respondents. Most nurses (63.6%) reported no need to stop or pause an infusion; pump malfunction was reported as a reason for interruption by 24.2%, and 12.1% reported slowing of the infusion. Among respondents to the respective questions, 96.9% reported frequently or almost always having more time for patient interaction, 81.3% reported less hand pain or discomfort, and 73.3% indicated that patients had reported less pain with mechanical pump-assisted administration compared with manual push. Based on ease of use, 97.0% of nurses indicated that they would recommend the mechanical system to other nurses using the manual push technique, while 100% would recommend it based on increased comfort.
Conclusion
Mechanical pump-assisted administration was well accepted by nurses and was associated with favorable nurse-reported experiences related to usability, patient interaction, hand comfort, and patient-reported pain compared with manual push administration. These findings suggest that a non-electronic mechanical infusion system may provide a practical alternative for short-duration subcutaneous administration. Prospective studies incorporating objective workflow measures and direct patient-reported outcomes are warranted to confirm these findings.

Introduction

Subcutaneous (SC) administration of medicinal products has expanded considerably in recent years, offering an alternative to intravenous (IV) administration across a growing range of therapeutic areas.1–3 For suitable therapies, the SC route can reduce or eliminate the need for IV access, shorten administration and chair time, and provide greater flexibility in where and how treatment is delivered.1,4 These advantages may improve the efficiency of healthcare delivery while reducing the burden associated with IV treatment for both patients and healthcare providers.1,5 As the number and diversity of medicinal products administered SC continue to increase, the practical aspects of administration, including the method and equipment used to deliver treatment, are becoming increasingly important considerations in clinical practice.6,7

Subcutaneous medicinal products may be administered using different delivery approaches, including manual push and pump-assisted administration, depending on factors such as the prescribed volume, product characteristics, administration time, clinical setting, and product-specific instructions.6–8 The choice of administration method also involves consideration of the overall delivery configuration, including the syringe, needle set, flow-control components, and infusion device, where applicable.7 Infusion pumps are well established for therapies requiring controlled or prolonged administration and can reduce the need for continuous hands-on administration by healthcare providers.6,9 However, for medicinal products with relatively short SC administration times, the potential value of pump-assisted delivery compared with manual push is less well characterized.8,10 In this setting, the administration method may influence not only the delivery itself, but also nursing workflow, healthcare provider ergonomics, patient interaction, and the overall treatment experience.10–12 For the purposes of this study, short-duration SC administration refers to treatments with administration times sufficiently short for manual push to represent a practical alternative to pump-assisted delivery.

Manual push administration requires the healthcare provider to remain actively engaged throughout the administration and to generate the pressure required to deliver the medicinal product from the syringe into the SC tissue.8,10 Depending on the volume and duration of the administration, maintaining manual pressure may contribute to hand discomfort or fatigue and may make it more difficult to maintain a consistent delivery rate.10 Continuous hands-on administration may also limit the healthcare provider's ability to perform other patient-care activities or interact freely with the patient during treatment.10,11

For short-duration SC administration, a mechanical infusion system represents an alternative to the continuous hands-on delivery required with manual push. A non-electronic, spring-driven mechanical infusion system uses controlled syringe pressure in combination with flow rate tubing to deliver SC medicinal products.9 By providing the driving force for administration, the system eliminates the need for the healthcare provider to continuously depress the syringe plunger and requires neither electronic programming nor an external power source.9 These characteristics may offer practical advantages for short-duration SC administration, where the simplicity of the delivery method must be balanced against nursing workload, ergonomics, and patient experience.9,11 However, evidence directly comparing healthcare providers' experience with mechanical pump-assisted administration and manual push for short-duration SC infusions remains limited.9,10

Against this background, the present study evaluated nurses' experience with a non-electronic, spring-driven mechanical infusion system compared with manual push for short-duration SC administration in routine hospital practice. The survey assessed ease of use and setup, training and troubleshooting, impact on nursing workflow and patient interaction, nurse-reported hand strain, perceived patient discomfort, and overall acceptance of the mechanical infusion system. The authors hypothesized that mechanical pump-assisted administration would be well accepted by nurses and could reduce the hands-on burden associated with manual administration while supporting healthcare provider-patient interaction and comfort.

Methods

Study design

This was a descriptive, cross-sectional survey study conducted among nurses involved in the administration of short-duration SC infusions in routine hospital practice in Denmark. The study was designed to characterize nurses' experiences with a non-electronic, spring-driven mechanical infusion system compared with their previous experience of manual push administration.

The participating hospitals had independently elected to use the mechanical infusion system for administration of the medicinal product in routine clinical practice. Prior to use, nurses received standard product training on the operation and use of the mechanical infusion system from the local distribution partner, in accordance with routine practice. After gaining practical experience with the infusion system through multiple SC administrations, nurses were invited to complete a structured survey evaluating their experience with the administration method. The survey focused on usability, training and troubleshooting, nursing workflow, healthcare provider ergonomics, perceived patient comfort, and overall acceptance of the mechanical system.

The study was descriptive in nature and was intended to capture healthcare providers’ experiences following use of the mechanical infusion system in routine clinical practice rather than to evaluate comparative clinical efficacy or establish causal relationships between administration methods.

Participants

Nurses were eligible to participate if they had direct experience routinely administering the SC medicinal product evaluated in this study using the manual push technique. Participants were specialist nurses involved in the regular administration of short-duration SC treatments in routine hospital practice.

Following standardized training on the operation of the infusion system, eligible nurses gained practical experience using the system before completing the survey. All eligible nurses at the participating sites were invited to participate, and participation was voluntary. No patient-level data were collected as part of the survey.

Questionnaire

A structured 12-item questionnaire was developed specifically for this study to evaluate nurses' experiences with mechanical pump-assisted administration and, where applicable, to compare these experiences with the manual push technique; it was not a previously validated instrument.

Usability was evaluated through items on overall ease of use, comfort with operating the pump, and clarity of the instructions provided (each on a 5-point scale, e.g., from "very difficult" to "very easy"), along with nurses' confidence in troubleshooting issues arising during administration (5-point scale from "not at all confident" to "very confident"). Setup efficiency was assessed by comparing pump setup time with manual push (somewhat longer, almost the same, or somewhat shorter), and an additional item captured whether an infusion required interruption and the type of issue encountered (leakage at the infusion site, twisted tubing, pump malfunction, or other).

The comparative experience of the two administration methods was further assessed with respect to patient-reported pain ("yes", "no", or "no difference"), the frequency of greater opportunity for nurse-patient interaction (5-point scale from "never" to "almost always"), and nurse hand pain or discomfort (5-point scale from "no, definitely not" to "yes, definitely"). Acceptability was measured by two items on willingness to recommend the pump to colleagues using the manual push technique, based on ease of use and on increased comfort (both on the same 5-point scale). The final item was an open question for suggestions for improving ease of use from predefined options (improved user interface, simplified instructions, clearer feedback during infusion, better troubleshooting resources, no improvements needed, or other), with an opportunity to provide an alternative response.

All outcomes were descriptive and based on nurses' self-reported experiences. Patient pain was assessed indirectly through nurses' reports of whether patients had reported less pain during pump-assisted administration; patients did not complete the questionnaire, and no patient-level clinical outcomes were collected.

Subcutaneous administration methods

Mechanical Pump-Assisted Administration. In the participating hospitals and therefore for this study, the FreedomEdge® Infusion System (KORU Medical Systems, Mahwah, NJ, USA) was used. It is a non-electronic, spring-driven mechanical syringe infusion system that uses controlled syringe pressure in combination with flow rate tubing to deliver medicinal products subcutaneously. The system does not require electronic programming, batteries, or an external power source. During administration, the system responds to changes in resistance at the SC infusion site, with the resulting flow rate decreasing as resistance increases.9,13,14

At the time of the survey, the medicinal product evaluated in this study was not included within the manufacturer’s indicated uses for the FreedomEdge® Infusion System. The decision to use the system for administration of this medicinal product was made independently by the participating hospitals as part of their routine clinical practice.

For mechanical pump-assisted administration, the following configuration was used (Figure 1): FreedomEdge® Infusion Pump (F10020); BD® Plastipak™ Luer-Lok® 20 mL syringe (Becton Dickinson & Co.; product number 300629); F600 Precision™ Flow Rate Tubing (F600); and a 24G, single-leg, 9 mm High-Flo™ SubQ Needle Set (RMS12409). The medicinal product (15 mL and 10 mL) was prepared in the 20 mL syringe and administered SC using the mechanical infusion system in accordance with the hospitals’ clinical procedures. The expected administration time with this configuration for 15 mL was approximately eight (8) minutes and five (5) minutes for 10 mL. The same administration configuration was used across participating hospitals.

Figure 1. Mechanical Infusion System (FreedomEdge®, KORU Medical Systems) and components used for subcutaneous administration.

Manual Push. For manual push administration, the medicinal product was prepared in a conventional syringe and administered subcutaneously by nurses by manually depressing the syringe plunger throughout the administration. The medicinal product, infusion volume, and intended administration time were the same as those used for mechanical pump-assisted administration. The specific syringe and SC needle configurations used for manual push administration have not been standardized and reflected routine clinical practice at the participating hospitals.

The principal difference between the administration approaches was the method used to generate the pressure required for delivery: manual pressure applied by the nurse during manual push administration versus spring-driven mechanical pressure provided by the infusion system during pump-assisted administration.

Training and clinical use

Prior to using the mechanical infusion system, nurses received standard product training from the manufacturer’s local distribution partner. Training covered preparation and setup of the infusion system, loading of the syringe, connection of the flow rate tubing and SC needle set, initiation and completion of the infusion, expected system operation, and basic troubleshooting.

Following training, nurses used the mechanical infusion system for administration of the medicinal product as part of routine clinical practice at their respective hospitals. Nurses gained practical experience with the system before completing the survey, allowing their responses to reflect experience with its use in clinical practice rather than training alone.

Data collection

The questionnaire was distributed through the nurses’ team managers, who also provided information about the study objectives and procedures before completing the questionnaire. The questionnaire was translated into the local language. It required approximately 30 minutes to complete. Thirty-three surveys were filled out by 33 nurses between February and March 2024. The survey was conducted in person at nine (9) hospitals in Denmark.

Statistical analysis

The analysis was conducted following completion of data collection. Quantitative data obtained from the closed-ended questions were analyzed using descriptive statistics. Categorical variables were summarized using absolute frequencies (n) and percentages (%).

No inferential statistical testing was performed because the primary objective of the study was to describe nurses’ preferences rather than to test predefined hypotheses or determine statistical associations between variables.

Responses to the open-ended questions were analyzed using a qualitative descriptive approach. All responses were reviewed and compared to identify recurring concepts and areas of similarity. Responses addressing similar issues were grouped into descriptive categories. The categories were subsequently reviewed and refined to ensure that they adequately represented the content of the responses. The frequency of occurrence of the identified categories was reported where appropriate. Representative anonymized responses were used to illustrate the main findings.

The findings from the open-ended questions were considered alongside the quantitative results to provide additional context and a more comprehensive description of nurses’ preferences.

Ethical considerations

The study was conducted in accordance with the ethical principles applicable to research involving human participants. All participants were informed about the purpose of the study, the voluntary nature of participation, and the handling of their data. Oral informed consent was obtained prior to participation. Data were collected anonymously and were stored and analyzed in accordance with applicable data protection requirements.

Results

A total of 33 nurses completed the survey. Overall, 90.9% (30/33) rated the mechanical infusion system as easy or very easy to use, while 9.1% (3/33) provided a neutral rating; none rated the system as difficult or very difficult to use. Most respondents (97.0%, 32/33) also reported feeling comfortable or very comfortable operating the system. Among nurses who responded to the question regarding clarity of the instructions for use (n=27), 81.5% (22/27) rated the instructions as clear or very clear, while 18.5% (5/27) provided a neutral rating; none rated the instructions as unclear.

When comparing setup time with manual push administration, 90.9% (30/33) reported that setup with the mechanical system was the same or shorter, while 9.1% (3/33) reported a longer setup time. During clinical use, 63.6% (21/33) reported no need to stop or pause an infusion. Pump malfunction was reported as a reason for stopping or pausing by 24.2% (8/33), while 12.1% (4/33) reported slowing of the infusion. Among respondents who answered the troubleshooting question (n=32), 59.4% (19/32) reported feeling confident or very confident in troubleshooting issues that might arise during an infusion, 37.5% (12/32) provided a neutral response, and 3.1% (1/32) reported not feeling confident.

Mechanical pump-assisted administration was associated with favorable nurse-reported experiences related to workflow and comfort. Among respondents to the patient-interaction question (n=32), 96.9% (31/32) reported that they frequently or almost always had more time to interact and communicate with patients when using the mechanical system, while 3.1% (1/32) reported occasionally having more time. No respondent reported an absence of increased patient interaction. Similarly, 81.3% (26/32) reported less hand pain or discomfort when using the mechanical system compared with manual push administration, while 15.6% (5/32) provided a neutral response and 3.1% (1/32) did not report less hand pain or discomfort.

Among nurses who responded to the question regarding patient pain (n=30), 73.3% (22/30) indicated that patients had reported less pain during mechanical pump-assisted administration compared with manual push administration. A further 20.0% (6/30) reported no difference in patient pain, while 6.7% (2/30) did not report less pain with mechanical pump-assisted administration.

Overall acceptance of the mechanical infusion system was high. Based on ease of use, 97.0% (32/33) of respondents indicated that they would probably or definitely recommend the system to other nurses using the manual push technique, while 3.0% (1/33) provided a neutral response. Based on increased comfort, all respondents (100%, 33/33) indicated that they would probably or definitely recommend the system.

When asked about potential improvements to enhance ease of use, 41.9% (13/31) indicated that no improvements were needed, 16.1% (5/31) suggested better troubleshooting resources, and 6.5% (2/31) suggested simplified instructions. The remaining 35.5% (11/31) selected “Other” and added comments like guiding symbols or other indicators on the pump to support the administration process. The survey results are presented in Table 1.

Table 1. Survey results

Survey domainResponse Optionsn (%)
Ease of use (n=33)Easy or very easy30 (90.9)
Neutral3 (9.1)
Difficult or very difficult0 (0.0)
Comfort operating the pump (n=33)Comfortable or very comfortable32 (97.0)
Neutral1 (3.0)
Uncomfortable or very uncomfortable0 (0.0)
Clarity of instructions for use (n=27)Clear or very clear22 (81.5)
Neutral5 (18.5)
Unclear or very unclear0 (0.0)
Set-up time vs manual push (n=33)Same or shorter30 (90.9)
Longer3 (9.1)
Stopping or pausing administration (n=33)No stoppage or pause21 (63.6)
Pump malfunction reported8 (24.2)
Slowing of administration4 (12.1)
Troubleshooting confidence (n=32)Confident or very confident19 (59.4)
Neutral12 (37.5)
Not confident1 (3.1)
Recommendation based on ease of use (n=33)Probably or definitely recommend32 (97.0)
Neutral1 (3.0)
Probably or definitely not recommend0 (0.0)
Recommendation based on increased comfort (n=33)Probably or definitely recommend33 (100.0)
Neutral0 (0.0)
Probably or definitely not recommend0 (0.0)
Patient interaction (n=32)Frequently or almost always more interaction31 (96.9)
Occasionally more interaction1 (3.1)
Rarely or never more interaction0 (0.0)
Patient-reported pain vs manual push (n=30)Patients reported less pain22 (73.3)
No difference6 (20.0)
Patients did not report less pain2 (6.7)
Nurse hand discomfort vs manual push (n=32)Less hand pain/discomfort26 (81.3)
Neutral5 (15.6)
Did not report less hand pain/discomfort1 (3.1)
Suggested improvements (n=31)No improvements needed13 (41.9)
Better troubleshooting resources5 (16.1)
Simplified instructions2 (6.5)
Other11 (35.5)

Data are presented as n (%). Denominators vary because not all respondents answered every survey item. Percentages are calculated using the number of respondents to each respective item.

Discussion

This multicenter, descriptive survey study found a high level of nurse acceptance of mechanical pump-assisted administration compared with routine manual push for short-duration SC infusions. Nurses generally reported favorable experiences with usability and comfort, and willingness to recommend the system was high based on both ease of use and increased comfort. Beyond usability, the findings identify potential benefits related to nurse-patient interaction, healthcare provider comfort, and patient experience.

One of the most notable findings was the reported increase in opportunity for nurse-patient interaction. Nearly all respondents to this question reported frequently or almost always having more time to interact and communicate with patients when using mechanical pump-assisted administration. Unlike manual push, which requires the nurse to continuously depress the syringe plunger throughout administration, the spring-driven system provides the mechanical force required for delivery without continuous manual involvement. This may allow nurses greater opportunity to communicate with patients or attend to other aspects of care during the administration. Although nursing time and healthcare utilization were not measured directly, the potential workflow implications of pump-assisted administration warrant further evaluation.

The findings also suggest a potential ergonomic benefit for nurses, with most respondents reporting less hand pain or discomfort than with manual push. Maintaining pressure on a syringe plunger may contribute to physical strain, particularly when multiple administrations are performed during routine clinical practice. Removing the need for continuous manual pressure may therefore offer an ergonomic advantage. However, hand discomfort was assessed through nurse self-report, and no objective ergonomic or workload measures were collected.

Patient experience was also perceived favorably. Almost three-quarters of respondents to the relevant question indicated that patients had reported less pain with mechanical pump-assisted administration than with manual push, while one-fifth reported no difference. This finding should be interpreted cautiously because patient pain was reported indirectly through nurses rather than measured directly using validated patient-reported measures of pain, comfort, and treatment preference.

From a practical usability perspective, most nurses reported that setup time was the same or shorter than with manual push. This is relevant because adding an infusion device might otherwise be expected to increase preparation burden for a relatively short administration. Together with the reduced need for continuous manual involvement, this may help explain the high level of acceptance observed in the study. However, actual setup and administration times were not prospectively measured; these findings therefore reflect nurses’ perceptions rather than objective time-and-motion data.

Most nurses reported no need to stop or pause administration, although interruptions attributed by respondents to pump malfunction and slowing of the infusion were reported. Slowing of the infusion may be consistent with the pressure-responsive operating characteristics of the mechanical system, whereby increasing resistance at the SC infusion site can result in a reduction in flow rate.9,13,14 Importantly, the survey did not include technical evaluation of individual events; consequently, events reported by nurses as pump malfunctions should not be interpreted as confirmed device malfunctions. Responses regarding troubleshooting confidence and suggestions for improved troubleshooting resources also indicate an opportunity to further support users through training and instructional materials.

Taken together, the findings suggest that the high acceptance of the mechanical pump-assisted administration reflects several aspects of the nursing experience rather than a single attribute. Ease of use and operating comfort were accompanied by perceived advantages related to patient interaction and nurse hand comfort, while indirect reports of patient pain suggest a potential patient-experience benefit. Free-text responses also identified opportunities to further improve usability, including symbols and indicators on the pump. These observations may inform both future evaluation and implementation of mechanical pump-assisted approaches for short-duration SC administration.

The present findings are consistent with previous evaluations of infusion systems for SC administration. In a head-to-head cross-sectional evaluation of two SC pumps used for immunoglobulin therapy, favorable setup times and patient preferences for mechanical pump-assisted delivery were reported11, and a structured literature review found comparable infusion-site reaction profiles between mechanical and electronic infusion systems9. Experiences with electronic infusion pumps, which require programming and an external power source, and with elastomeric balloon pumps, which deliver at fixed flow rates determined by the device, suggest that device characteristics such as setup complexity, visibility of infusion progress, and flexibility of flow-rate selection strongly influence nurse and patient acceptance.9,11 The high ease-of-use ratings observed in the present study suggest that the simplicity of a non-electronic, spring-driven system may be advantageousin the short-duration administration setting. Similarly, a qualitative study of SC oncology monoclonal antibody delivery identified nursing workload and administration practicalities as key challenges,10 which aligns with the workflow and interaction benefits reported by nurses in the present study.

Beyond these device comparisons, the ergonomic burden of manual push administration is increasingly documented. In the same qualitative study, ergonomic difficulties during manual push were reported by 5 of 9 nurses, who described physical discomfort and being “tethered” to patients throughout administration, and concerns have been raised regarding the risk of nurse musculoskeletal disorders caused by repeated manual syringe injections.10 A recent review of SC delivery devices for oncology medicinal products similarly noted that manually pushing a syringe is labor-intensive for nurses and requires continuous chairside presence.15 More direct comparative strain data are available from a quality improvement pilot at four community infusion centers, reported in poster form: across 64 documented SC monoclonal antibody injections, mean physical strain ratings decreased from 3.0 to 1.0 on a 10-point scale after implementation of a mechanical infusion pump, with all paired ratings equal to or lower than with previous manual push; nurses reported being able to multitask during 80% of pump-assisted infusions, and patient-reported pain and satisfaction were favorable.16 Although that pilot was small and conducted in community oncology settings, its findings align closely with the reduced hand discomfort and increased opportunity for patient interaction reported by nurses in the present study.

Although the present study did not collect economic endpoints, the workflow-related findings may have economic implications. Systematic reviews of SC versus IV administration of oncology biologics have reported substantial reductions in chair time (approximately 55−77 minutes) and healthcare provider time (approximately 13−68 minutes) per administration, with associated cost savings.17,18 In time-and-motion studies of SC trastuzumab, active healthcare provider time was reduced primarily because the infusion, once initiated, did not require continuous hands-on involvement.19 By analogy, mechanical pump-assisted SC administration in the present setting may release nursing time for other patient-care activities, and the reported setup times comparable to manual push suggest that such potential time savings are not offset by added preparation burden. Formal economic evaluations incorporating objective time measurements and cost analyses would be required to quantify these potential benefits, and the present descriptive findings should be considered hypothesis-generating in this regard.

Limitations

The study was descriptive, included a relatively small sample of 33 nurses, and used a non-validated, study-specific questionnaire based on self-reported experience rather than objective comparative measures. The survey did not include a randomized or controlled comparison between administration methods. These limitations may affect the generalizability of the findings and preclude conclusions regarding comparative clinical effectiveness.

Conclusions

This multicenter survey study found that the mechanical infusion system was well accepted by nurses as an alternative to manual push for short-duration SC administration. Most nurses rated the system as easy to use and comfortable to operate and indicated that they would recommend it to other nurses using the manual push technique.

The findings suggest that mechanical pump-assisted administration may provide a practical and user-friendly alternative to manual push for short-duration SC infusions. Further prospective research incorporating objective workflow measures, standardized administration procedures, and direct patient-reported outcomes is warranted to confirm these findings and better characterize the potential benefits for patients, healthcare providers, and clinical practice.

Contribution statements

  • This study provides multicenter real-world evidence on nurses’ experiences with a non-electronic mechanical infusion system as an alternative to manual push for short-duration subcutaneous administration.
  • Mechanical pump-assisted administration was associated with favorable nurse-reported experiences related to usability, patient interaction, healthcare provider hand comfort, and patient-reported pain.
  • The findings identify potential workflow, ergonomic, and patient-experience benefits that warrant confirmation in prospective studies using objective measures and direct patient-reported outcomes.

Declarations

Declaration of Interests
Anna Majapuro-Hirvonen, Brent Rutland, and Jasmin Bosshard are employees of KORU Medical Systems, the manufacturer of the FreedomEdge® Infusion System evaluated in this study. The authors declare no other competing interests.
The participating hospitals independently elected to use the FreedomEdge® Infusion System for administration of the medicinal product in routine clinical practice. Product training was provided by KORU Medical Systems’ local distribution partner as part of its routine training activities. KORU Medical Systems did not direct the hospitals’ decision to use the infusion system.
Declaration of Funding
No external funding was received for this work. KORU Medical Systems provided nominal compensation to participating hospitals for completed nurse surveys. No financial support was provided for the hospitals’ clinical use of the infusion system.
Acknowledgements
The authors would like to thank the nurses at the participating hospitals in Denmark for their time, participation, and valuable contributions to this study.

References

  1. Bittner B, Richter W, Schmidt J. Subcutaneous Administration of Biotherapeutics: An Overview of Current Challenges and Opportunities. BioDrugs. 2018;32(5):425–440. PMID: 30043229; PMCID: PMC6182494.
  2. Usach I, Martinez R, Festini T, Peris J-E. Subcutaneous Injection of Drugs: Literature Review of Factors Influencing Pain Sensation at the Injection Site. Adv Ther. 2019;36(11):2986–2996. PMID: 31587143; PMCID: PMC6822791.
  3. Broadhurst D, Cooke M, Sriram D, Gray B. Subcutaneous hydration and medications infusions (effectiveness, safety, acceptability): A systematic review of systematic reviews. PLoS ONE. 2020;15(8):e0237572. PMID: 32788803; PMCID: PMC7427522.
  4. Tahami Monfared AA, et al. Societal Costs and Efficiency of Subcutaneous versus Intravenous Lecanemab in Early Alzheimer's Disease. Neurol Ther. 2025;14(5):1863–1888. PMID: 40610846; PMCID: PMC12450161.
  5. Skoda-Smith S, Torgerson TR, Ochs HD. Subcutaneous immunoglobulin replacement therapy in the treatment of patients with primary immunodeficiency disease. Ther Clin Risk Manag. 2010;6:1–10. PMID: 20169031; PMCID: PMC2817783.
  6. Green P, Schneider A, Lange J. Navigating large-volume subcutaneous injections of biopharmaceuticals: a systematic review of clinical pipelines and approved products. mAbs. 2024. PMID: 39279181; PMCID: PMC11407384.
  7. Grosse-Kreul D, Allen C, Kalicinsky C, Keith PK. Infusion parameters, safety, and practical guidance for the manual administration of subcutaneous immunoglobulin 20% (Ig20Gly). Allergy Asthma Clin Immunol. 2024. PMID: 39367472; PMCID: PMC11453067.
  8. Cowan J, et al. Safety and Tolerability of Manual Push Administration of Subcutaneous IgPro20 at High Infusion Rates (HILO Study). J Clin Immunol. 2020. PMID: 33025378; PMCID: PMC7846525.
  9. Majapuro-Hirvonen A, Rutland B, Bosshard J. Mechanical versus Electronic Infusion Systems in Subcutaneous Immunoglobulin Therapy: A Structured Literature Review of Infusion-Site Reactions and Device Performance. Med Devices Evid Res. 2026. PMID: 42540737; PMCID: PMC13426497.
  10. Franzese C, Anderson M, Wu J, et al. Current practices, challenges, and future directions in subcutaneous oncology monoclonal antibody delivery: A qualitative study. medRxiv preprint, 2026. https://www.medrxiv.org/content/10.64898/2026.05.06.26352582v1.full-text
  11. Rutland B, Strasters S, Bosshard J. Comparing setup time and patient preferences for SCIg therapy: a head-to-head cross-sectional evaluation of two SC pumps. 2025. PMID: 40530995; PMCID: PMC12184102.
  12. Zheng R, et al. Reducing treatment burden in HER2+ breast cancer. 2026. PMCID: PMC12851754.
  13. KORU Medical Systems. FreedomEdge® Syringe Infusion System Instructions for Use. Document 347201, Rev A. Mahwah, NJ: KORU Medical Systems; 2022.
  14. Pratt CJ, Rutland B, Southworth C, Wall D. Enhancing Efficiency in Home-Based Subcutaneous Immunoglobulin Therapy: A Comparative Analysis of Supply Selection for Infusion Time Optimization. Int J Nurs Health Care Res. 2025;8:1657. doi:10.29011/2688-9501.101657.
  15. Bobin A, Ocio EM, Parmar G, et al. Review of the first on-body injector (OBI) for the subcutaneous delivery of an oncology drug: isatuximab in multiple myeloma. Expert Opin Drug Deliv. 2026. doi:10.1080/17425247.2026.2659263
  16. Wall D, McIntyre PJ, Fano R, Wardell J, Morefield E, Wojcik E. Reducing Ergonomic Strain During Subcutaneous Monoclonal Antibody Administration. Poster presentation; 2026. https://www.ovid.com/journals/ocnf/abstract/00006549-202603000-00758~reducing-ergonomic-strain-during-subcutaneous-monoclonal?redirectionsource=fulltextview
  17. McCloskey C, Ortega MT, Nair S, Garcia MJ, Manevy F. A systematic review of time and resource use costs of subcutaneous versus intravenous administration of oncology biologics in a hospital setting. Pharmacoecon Open. 2022;7(1):3-36. doi:10.1007/s41669-022-00361-3. PMID: 35996066
  18. George S, Bourlon MT, Overman MJ, et al. Systematic literature review of intravenous versus subcutaneous administration of oncology therapies: a clinical, economic and patient perspective. Cancer Treat Rev. 2025;139:102974. doi:10.1016/j.ctrv.2025.102974
  19. De Cock E, Pivot X, Hauser N, et al. A time and motion study of subcutaneous versus intravenous trastuzumab in patients with HER2-positive early breast cancer. Cancer Med. 2016;5(3):389-397. doi:10.1002/cam4.573. PMID: 26806010

Article information

Authors

How to cite

Majapuro-Hirvonen, A., Kallehauge Terndrup, H., et al., 2026. Mechanical Pump-Assisted versus Manual Push Administration for Short-Duration Subcutaneous Infusions: A Nurses’ Preference Study. Medical Research Archives, [online] 14(10). https://doi.org/10.18103/mra.2026.0617